While the device is designed for automation, the "manual" aspect of operation covers daily start-up/shutdown procedures, manual sample dilution, manual maintenance modes, and troubleshooting flagged results.
— Rachel M., Laboratory Manager
The Advia 560!
| Error Message | Probable Cause | User Action | |---------------|----------------|--------------| | | Debris in WBC aperture | Run Clean Aperture . If persists, perform manual cleaning. | | HGB Too Low | Lyse empty or air bubble | Check lyse reagent; prime the lyse line. | | Sample Clot | Microclot in tube | Remix sample; use a new tube. Run Clot Removal . | | Vacuum Error | Waste container full or leak | Empty waste; check waste tubing connections. | | Diluent Empty | Diluent container dry | Replace diluent; run Prime > Diluent . | | No Scattergram | Flow cell dirty or laser failed | Clean flow cell (monthly maintenance). If fails, call service. | advia 560 manual